logo
  1. Home
  2. Sonda termiczna, Ø 10 mm, możliwość 90° wygięcia (200 mm), ...

Sonda termiczna, Ø 10 mm, możliwość 90° wygięcia (200 mm), ... - for flow, temperature, humidity and pressure

0635 1543
  • 4 functions in 1 probe: measurement of flow, temperature, humidity and absolute pressure
  • With robust, non-rotatable telescope (can be extended to 1100 mm)
  • Intelligent calibration concept
  • Flow measuring range: 0 to +20 m/s
  • Produkt nie jest już dostępny. Czy znają Państwo już kolejny model?
    Warto skorzystać z jego zalet:
    >> Do produktu(wersja przewodowa)
    >> Do produktu (wersja bezprzewodowa, z funkcją Bluetooth)
    Prosimy pamiętać, że kolejne modele są kompatybilne wyłącznie z testo 400 lub testo 440.
    Use the hot wire probe (thermal anemometer) to determine flow velocity and volume flow in ventilation ducts or at wall/ceiling outlets. It is particularly suitable for low and medium flow velocities up to +20 m/s. The probe can also measure temperature, humidity and absolute pressure simultaneously.

    The hot wire probe is equipped with a robust, non-rotatable telescope, which can be extended to a maximum length of 1100 mm. The immersion depth is easy to read off thanks to the scaling on the telescope.

    For easier handling, the probe can be bent up to an angle of 90°. On the handle there is a practical measuring button, which activates storage of the readings. So you can fully focus on the positioning of the probe in the ventilation duct.

    Intelligent calibration concept

    The probe offers maximum digital measurement reliability. The digital probe allows readings to be processed directly in the probe. This technology eliminates instrument measurement uncertainty. The probe can be returned on its own (without the measuring instrument) for calibration. Calculating the determined calibration data in the probe generates a zero-error display.

    Hot wire probe (Ø 10 mm) with telescope (can be extended to 1100 mm).

    Note: you need a plug-in head cable (0430 0100) for this probe.

    Pomiar temperatury - NTC

    Zakres pomiarowy
    -20 do +70 °C
    Dokładność
    ±0,5 °C

    Wilgotność - czujnik pojemnościowy

    Zakres pomiarowy
    0 do 100 %RH
    Dokładność
    ±(1,8 %RH + 0,7 % mierz.wart.)

    Ciśnienie absolutne

    Zakres pomiarowy
    +700 do +1100 hPa
    Dokładność
    ±3 hPa