testo 330-2 LL

Order-Nr.  0563 3372 77

Flue gas analyzer with longlife gas sensors and integrated draught and gas zeroing


Details

Product Description

Thanks to the new instrument functions, the Testo flue gas analyzers of the new testo 330-LL series now offer you even more professional and reliable support. These are the main features:
• High-resolution colour display for the graphic representation of your measurement data
• Extended measurement menus – such as gas pipe testing – for a comprehensive analysis of your heating system
• Logger function for easy long-term recording of the measurement curve

Technical Data

General technical data

Weight

600 g (without rechargeable battery)

Dimensions

270 x 90 x 65 mm

Operating temperature

-5 to +45 °C

Display size

240 x 320 pixels

Display function

Colour graphic display

Power supply

Rechargeable battery pack 3.7 V / 2.6 Ah; Mains unit 6 V / 1.2 A

Maximum memory

500,000 readings

Storage temperature

-20 to +50 °C

Differential Pressure - Piezoresistive

Measuring range

±10000 Pa

Accuracy

±0.3 Pa (0 to 9.99 Pa) plus ±1 Digit

±3 % of mv (10 to 10000 Pa) plus ±1 Digit

Flue gas O₂

Measuring range

0 to 21 Vol.%

Accuracy

±0.2 Vol.%

Resolution

0.1 Vol.%

Reaction time t₉₀

< 20 s

Flue gas CO (with H₂-compensation)

Measuring range

0 to 8000 ppm

Accuracy

±10 ppm or ±10 % of mv (0 to 200 ppm)

±20 ppm or ±5 % of mv (201 to 2000 ppm)

±10 % of mv (2001 to 8000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 60 s

Flue gas COlow (with H₂-compensation)

Measuring range

0 to 500 ppm

Accuracy

±2 ppm (0 to 39.9 ppm)

±5 % of mv (40 to 500 ppm)

Resolution

0.1 ppm

Reaction time t₉₀

< 40 s

CO determination (with H₂-compensation), automatic dilution

Measuring range

0 to 30000 ppm

Accuracy

±100 ppm (0 to 1000 ppm)

±10 % of mv (1001 to 30000 ppm)

Resolution

1 ppm

Flue gas NO

Measuring range

0 to 3000 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5 % of mv (101 to 2000 ppm)

±10 % of mv (2001 to 3000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 30 s

Flue gas Nolow

Measuring range

0 to 300 ppm

Accuracy

±2 ppm (0 to 39.9 ppm)

±5 % of mv (40 to 300 ppm)

Resolution

0.1 ppm

Reaction time t₉₀

< 30 s

Differential Pressure flue gas Draught - Piezoresistive

Measuring range

-9.99 to +40 hPa

Accuracy

±0.02 hPa or ±5 % of mv (-0.50 to +0.60 hPa)

±0.03 hPa (+0.61 to +3.00 hPa)

±1.5 % of mv (+3.01 to +40.00 hPa)

Resolution

0.01 hPa

Temperature

Measuring range

-40 to +1200 °C

Accuracy

±0.5 °C (0 to +100.0 °C)

±0.5 % of mv (Remaining Range)

Resolution

0.1 °C (-40 to +999.9 °C)

1 °C (> +1000 °C)

Flue gas degree of effectivity, Eta (calculated)

Measuring range

0 to 120 %

Resolution

0.1 %

Flue gas loss (calculated)

Measuring range

0 to 99.9 %

Resolution

0.1 %

Flue gas CO₂ calculation (calculated from O₂)

Measuring range

0 to CO₂ max (Display range)

Accuracy

±0.2 Vol.%

Resolution

0.1 Vol.%

Reaction time t₉₀

< 40 s

Velocity / Volume flow

Measuring range

0.15 to 3 m/s

Resolution

0.1 m/s

Pressure measurement

Measuring range

0 to +300 hPa

Accuracy

±0.5 hPa (0.0 to 50.0 hPa)

±1 % of mv (50.1 to 100.0 hPa)

±1.5 % of mv (Remaining Range)

Resolution

0.1 hPa

Flue gas CO (without H₂-compensation)

Measuring range

0 to 4000 ppm

n1{#0} t1 n2{#0} {%S2}

Accuracy

±20 ppm (0 to 400 ppm)

±5 % of mv (401 to 2000 ppm)

±10 % of mv (2001 to 4000 ppm)

Resolution

1 ppm

Reaction time t₉₀

< 60 s

Ambient CO

Measuring range

0 to 500 ppm

Accuracy

±5 ppm (0 to 100 ppm)

±5 % of mv (> 100 ppm)

Resolution

1 ppm

Reaction time

Approx. 35 s

with CO probe

Ambient CO₂

Measuring range

0 to 1 Vol.%

0 to 10000 ppm

Accuracy

±50 ppm or ±2 % of mv (0 to 5000 ppm)

±100 ppm or ±3 % of mv (5001 to 10000 ppm)

Reaction time

Approx. 35 s

with ambient CO₂ probe

Gas leak measurement for combustible gases (via gas leak detection probe)

Measuring range

0 to 10000 ppm CH₄ / C₃H₈; Display range

Accuracy

Signal optical display (LED) audible signal via buzzer

Reaction time t₉₀

< 2 s

with gas leak detection probe

Temperature (via the fine pressure probe)

Measuring range

-40 to +1200 °C max. (dependent on probe)

Accuracy

±0.5 °C (-40 to 100 °C)

±0.5 % of mv (Remaining Range) plus probe accuracy

Resolution

0.1 °C




Accessories

Accessories for probes

AED 1.361
excl. VAT
AED 1.889
excl. VAT
AED 599
excl. VAT

Measurement cells

O2 sensor for testo 330-1 LL/-2 LL

Order Number: 0393 0002

AED 3.030
excl. VAT

Downloads

Software / Firmware

  • Instruction manual EasyHeat software  (pdf, 3.55 MB)

  • Testo ZIV driver ZIV 2000 for testo 320 and testo 330  (v2.1, 2.22 MB) Testo ZIV driver in the 2000 version. The Testo ZIV driver is used to connect the testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface Version 2.0 defined by the Zentralverband des Schornsteinfegerhandwerks (Central Association of Chimney Sweeps, ZIV). Please check with the manufacturer of your application program as to whether this interface is supported. If Microsoft .NET Framework 4.0 has not been installed on the computer, it must be downloaded from the Microsoft website and installed on the system.

  • Testo ZIV driver for testo 300, testo 320 and testo 330  (v2.3, 64.11 MB) The Testo ZIV driver is used to connect the testo 300, testo 320 and testo 330 measuring instruments to an application program (sweeping district administration program) according to the interface defined by the Zentralverband des Schornsteinfegerhandwerks (ZIV, Central Association of Chimney Sweeps) in version 1.0 of 01. August 2012, in version 2.0 of 13. February 2017 as well as version 3.0 from 02. July 2021. Please check with the manufacturer of your application program as to whether this interface is supported.

  • Update-Kit / Bootloader  (V1.22, 1.24 MB) (testo 330 LL | testo 330i | testo 350 Control Unit + Analysis Box | testo 320)
    If the firmware update does not start under Windows 8.1 or Windows 10, a new bootloader must be installed on the measuring device once.